Search for...

Or try a general search...

    CLOSE

    Model 410

    ← Back to products

    Product Overview

    All three Models of 410 Flame Photometer come with a preloaded filter stick which houses a maximum of three element specific filters. The filter stick is easily removed and replaced within the chimney housing (no tools required) and as such it is a relatively simple operation to introduce alternative element specific filters.

    Filters available for use with all Model 410’s include: Sodium, Potassium, Lithium, Calcium, Barium, Strontium, Rubidium and Caesium; the most extensive range available from any Flame Photometer Manufacturer.

    The Model 410 Classic Flame Photometer (475 41 200) is the one developed by Corning over 30 years ago. We still manufacture this highly successful unit due to popular demand.  The Model 410 Classic Flame Photometer has Sodium, Potassium and Calcium specific filters and NIST traceable 1000 ppm standard concentrates for each of those analytes (6 x 100 ml). We have expanded the Model 410 Classic’s utility by introduction of the:

    Model 410 Industrial and Model 410C Flame Photometers; both come with a Digital Interface fitted as standard.  The Digital Interface allows connection to a serial printer or computer for result collection.

    Accessories

    M410 BlueNotes software (410 66 000)

    M410 BlueNotes  is a software package designed not just for data collection, storage, manipulation and report generation; it can also improve throughput, enhance precision and extend the utility of the Model 410 Flame Photometer.

    Everything is visible on one operational page

    With the M410 BlueNotes software package you can;

    • Use real names and sample numbers in your reports
    • Perform Multipoint calibrations with curve fit
    • Automatically correct for instrument drift
    • Save calibration curves
    • Save methods
    • Archive results
    • Prepare and print professional reports
    • Use Automatic Peak Selection facility to assess stability of readings
    • Automate analyses with addition of the Model 860 Autosampler

    Model 410 Flame Photometer Instrument Update and Automation items and packages

    410 86 001* Digital Interface/Linearisation Module

    410 86 002* BlueNotes Upgrade (410 86 001 & 410 66 000)

    410 86 003* BlueNotes Upgrade & Automation Package (860 00 009 M860 Autosampler & 410 86 002)

    410 66 000* M410 BlueNotes Software

    410 86 004* BlueNotes Automation Package (860 00 009 & 410 66 000)

    *When ordering any of the above products/packages, you need to state the serial number of the instrument with which you intend to use them.

    The Model 410 Industrial Flame Photometer (475 41 201)

    Flame Photometry is the technique of choice for the measurement of Sodium, Potassium and Calcium in all sample types in: Mineral Extraction, Oil Industry, Paper Industry, Pharmaceuticals, Soil Analysis, Utilities, Food & Beverage, Chemical Manufacture and Fertilisers. The Model 410 Industrial Flame Photometer with digital output, for printer or PC connection, also has Sodium linearisation capability to allow higher concentrations of Sodium (to 20 ppm) to be directly measured using a single point calibration. That facility can be switched on or off according to the operator’s preference or protocol for Sodium determinations. The Model 410 Industrial Flame Photometer has Sodium, Potassium and Calcium specific filters and NIST traceable 1000 ppm standard concentrates for each of those analytes (6 x 100 ml).

    Model 410C Flame Photometer (475 41 301)

    For biological samples, Flame Photometry is the most sensitive and robust method for the determination of Sodium, Potassium and Lithium. The Model 410C Flame Photometer is delivered with Sodium, Potassium and Lithium filters pre-fitted and an appropriate multi-element calibrator in mmol/l. (1 x 100ml). The Model 410C instrument is also fitted with a Digital Interface and Sodium linearisation facility allowing direct, straight line, calibration of clinically significant concentrations of Sodium, as well as Potassium and Lithium, following suitable dilution. (Sherwood’s Model 805 Continuous Flow Diluter, used together with the Model 410C Flame Photometer, gives nominal sample dilution ratios of 1:200 and 1:50. It has been designed to give consistent dilution of blank, calibrator and samples to the Flame Photometer).

    Both Model 410 Industrial and Model 410C Flame Photometers can be further enhanced by the use of Sherwood’s Model 410 BlueNotes software (410 66 000) and Model 860 Autosampler (860 00 009). Fully integrated and automated systems are possible, even for the Model 410 Classic Flame Photometer via our Update and Automate Modules and packages.

     

     

    Model 410 Classic Specification

    OPERATING RANGE

    Sodium is highly prone to self- absorbance in the flame above a concentration of 10 ppm. The Digital Interface/Lineariser Module accessory (see above) will extend the linear response to 20ppm.

    RANGE Displayed

    Na 0 – 1999

    K 0 – 1999

    Ca 0 – 1999

    MINIMUM DETECTION LEVEL

    @3 x noise

    SAMPLE CONCENTRATION

    to give min. 100 Units and still be linear (not Sr)

    LINEARITY

    % of full value at 50% of value

    ppm

    mg/l

    meq/l

    ppm

    mg/l

    meq/l

    Na

    0.02

    0.02

    0.0009

    0.5 –10

    0.5 –10

    0.02 – 0.45

    ± 1% @ 3 ppm

    K

    0.02

    0.02

    0.0005

    0.5 – 20

    0.5 – 20

    0.01 – 0.5

    ± 1% @ 3 ppm

    Ca

    0.2

    0.2

    0.003

    10 – 100

    10 – 100

    0.125 – 1.25

    ± 1% @ 100 ppm

    Li

    0.02

    0.02

    0.003

    0.5 – 20

    0.5 – 20

    0.07 – 2.88

    ± 1% @ 3 ppm

    Ba

    20

    20

    0.07

    1000

    1000

    3.65

    ± 1% @ 1000 ppm

    Cs

    0.2

    0.2

    0.002

    20 – 100

    20 – 100

    0.15 – 0.75

    ± 5% @ 100 ppm*

    Rb

    0.1

    0.1

    0.001

    1 – 100

    1 – 100

    0.01 – 1.17

    ± 1% @ 100 ppm

    Sr

    20

    20

    0.1

    1000

    1000

    5.71

    Nonlinear > 200 ppm

    * for best results construct a calibration curve for the concentration range of Cs of interest.

    SPECIFICITY

    Na, K & Li Interference is < 0.5% value of element under analysis.

    REPRODUCIBILITY  (K=10ppm 20 aliquots from single sample)

    < 1.5%

    ASPIRATION RATE

    3-6 ml/min

    SPEED OF ANALYSIS

    Time to stability 15 seconds

    DATA OUTPUT

    Analogue: Voltage 0.20 Volts FSD

    DIMENSIONS

    510mm x 390mm x 345mm 9.5 Kg Net

    730mm x 450mm x 450mm 17 Kg Packed

     

     

     

    Model 410 Industrial Specification

    #OPERATING RANGE

    Sodium is highly prone to self- absorbance in the flame above a concentration of 10 ppm.

    The Printer Lineariser Module accessory fitted to the Model 410 Industrial extends the linear

    response to 20 ppm.

    RANGE Displayed

    Na 0 – 1999

    K 0 – 1999

    Ca 0 – 1999

    MINIMUM DETECTION LEVEL

    @3 x noise

    SAMPLE CONCENTRATION

    to give min. 100 Units and still be linear

    LINEARITY

    % of Nominal value when measured at 50% Nominal value

    Na

    20 ppb

    0.5 – 10 ppm **

    ± 1% @ 3 ppm

    K

    20 ppb

    0.5 – 20 ppm

    ± 1% @ 3 ppm

    Ca

    0.2 ppm

    10 – 100 ppm

    ± 1% @ 100 ppm

    Li

    20 ppb

    0.5 – 20 ppm

    ± 1% @ 3 ppm

    Ba

    20 ppm

    1000 ppm

    ± 1% @ 1000 ppm

    Cs

    0.2 ppm

    20 – 100 ppm

    ± 5% @ 100 ppm*

    Rb

    0.1 ppm

    1 – 100 ppm

    ± 1% @ 100 ppm

    Sr

    20 ppm

    1000 ppm

    Non linear > 200 ppm

    * for best results construct a calibration curve for the concentration range of Cs of interest.

    ** 20 ppm with Lineariser switched on.

    SPECIFICITY

    Na, K & Li Interference is < 0.5% value of element under analysis.

    REPRODUCIBILITY  (K=10 ppm 20 aliquots from single sample)

    < 1.5%

    ASPIRATION RATE

    3-6 ml/min

    SPEED OF ANALYSIS

    Time to stability 15 seconds

    DATA OUTPUT

    Analogue: Voltage 0.20 Volts FSD

    DIMENSIONS

    510mm x 390mm x 345mm 9.5 Kg Net

    730mm x 450mm x 450mm 17 Kg Packed

     

    Model 410C Specification

    OPERATING RANGE

    Sodium is highly prone to self- absorbance in the flame above a concentration of 10 ppm.

    A Lineariser is fitted to the Model 410C for clinical applications where normal sodium concentrations

    even after dilution  at 200:1 exceed this concentration.

    Potassium and Lithium do not exhibit self- absorbance.

    RANGE Displayed

    Na 0 – 1999

    K 0 – 1999

    Li 0 – 1999

    REPRODUCIBILITY  (K=10ppm 20 aliquots from single sample)

    < 2%

    SAMPLE CONCENTRATION RANGE

    (before dilution)

    NB Dilution for

    Clinical Samples

    Linearity

    Reproducibility

    Cv

    Na   Serum  110 – 170 mmol/l

    1:200

    ± 2 mmol/l

    < 1.5%

    Na   Urine       0 – 200 mmol/l

    1:200

    ± 4 mmol/l

    < 1.5%

    K    Serum      0 – 10.0 mmol/l

    1:200

    ± 0.2 mmol/l

    < 1.5%

    K   Urine         0 – 200 mmol/l

    1:5 then 1:200

    ± 2.0 mmol/l

    < 1.5%

    Li                     0 – 3.0 mmol/l

    1:50

    ± 0.2 mmol/l

    < 2.5%

    SPECIFICITY

    Na, K & Li Interference is < 1.0 mmol/l with normal levels of other elements in Serum/urine.

    ASPIRATION RATE

    3-6 ml/min

    SPEED OF ANALYSIS

    Time to stability 15 seconds

    DATA OUTPUT

    Analogue: Voltage 0.20 Volts FSD

    DIMENSIONS

    510mm x 390mm x 345mm 9.5 Kg Net

    730mm x 450mm x 450mm 17 Kg Packed